EKG artifacts

Jump to navigation Jump to search

WikiDoc Resources for EKG artifacts

Articles

Most recent articles on EKG artifacts

Most cited articles on EKG artifacts

Review articles on EKG artifacts

Articles on EKG artifacts in N Eng J Med, Lancet, BMJ

Media

Powerpoint slides on EKG artifacts

Images of EKG artifacts

Photos of EKG artifacts

Podcasts & MP3s on EKG artifacts

Videos on EKG artifacts

Evidence Based Medicine

Cochrane Collaboration on EKG artifacts

Bandolier on EKG artifacts

TRIP on EKG artifacts

Clinical Trials

Ongoing Trials on EKG artifacts at Clinical Trials.gov

Trial results on EKG artifacts

Clinical Trials on EKG artifacts at Google

Guidelines / Policies / Govt

US National Guidelines Clearinghouse on EKG artifacts

NICE Guidance on EKG artifacts

NHS PRODIGY Guidance

FDA on EKG artifacts

CDC on EKG artifacts

Books

Books on EKG artifacts

News

EKG artifacts in the news

Be alerted to news on EKG artifacts

News trends on EKG artifacts

Commentary

Blogs on EKG artifacts

Definitions

Definitions of EKG artifacts

Patient Resources / Community

Patient resources on EKG artifacts

Discussion groups on EKG artifacts

Patient Handouts on EKG artifacts

Directions to Hospitals Treating EKG artifacts

Risk calculators and risk factors for EKG artifacts

Healthcare Provider Resources

Symptoms of EKG artifacts

Causes & Risk Factors for EKG artifacts

Diagnostic studies for EKG artifacts

Treatment of EKG artifacts

Continuing Medical Education (CME)

CME Programs on EKG artifacts

International

EKG artifacts en Espanol

EKG artifacts en Francais

Business

EKG artifacts in the Marketplace

Patents on EKG artifacts

Experimental / Informatics

List of terms related to EKG artifacts

Editor-in-Chief: C. Michael Gibson, M.D.; Santosh Patel M.D., FRCA [1]; Associate Editor-In-Chief: Kashish Goel, M.D.; Cafer Zorkun, M.D., Ph.D. [2]

Overview

Electrocardiograph (EKG) artifacts are defined as EKG abnormalities, which are a measurement of cardiac potentials on the body surface and are not related to electrical activity of the heart.[1] As a result of artifacts, normal components of the EKG can be distorted.[2] It is very important to recognize these artifacts, otherwise they can lead to unnecessary testing and therapeutic interventions. In this chapter, we will present the common causes and ways to characterize EKG artifacts.

Causes

EKG artifacts can be generated by internal and external causes[3].

Internal

These are physiological causes that could be due to:

  • Patient's motion: Does not allow electronic filtration (large swings, usually by epidermal stretching).
    • Tremors and shivering cause motion artifacts.
    • Simple movements such as brushing and combing the hair can produce EKG disturbances during ambulatory EKG monitoring.
  • Muscular activity: Allows electronic filtration (small spikes).[3]

External

These are non-physiological causes associated with other electrical devices attached to or implanted (e.g. deep brain stimulator) in the body and includes the following[3]:

  • Electromagnetic interference:
    • Power line electrical disturbances/ Light fixtures
    • Electrocautery
    • Electrical devices in the room
    • Radiofrequency based commercial (e.g. mobile phones) products
  • Cable and electrode malfunction:
    • Insufficient electrode gel
    • Misplaced leads
    • Inappropriate filter settings
    • Broken wires
    • Loose connections
    • Accumulation of static energy
  • Medical equipments: In operation theatres and intensive care unit various equipments can affect EKG monitoring system (e.g. electrodes, leads, amplifier, filters).[4][5][6]

Medical equipment related EKG artifacts

Type of equipment Artifact
IVAC intravenous infusion controller Atrial or ventricular extrasystoles, pseudowaves (QRS)
Cardiopulmonary bypass pump Uninterpretable tracing, non-specific
Pressure-controlled irrigation pump Atrial flutter
COBE Prisma System for continuous venovenous hemofiltration Atrial flutter
Flexible bronchoscope Atrial fibrillation
Deep brain stimulator Uninterpretable tracing
Straight shot microdebrider (nasal endoscopy) Ventricular tachycardia
Intra-aortic balloon pump Pseudowaves (P), premature atrial contraction
Somatosensory evoked potential monitoring units Supraventricular tachycardia
High-frequency oscillatory ventilation Atrial flutter, atrial fibrillation, ventricular tachycardia
Intraoperative high-field MRI Ventricular tachycardia, ventricular fibrillation, non-specific
Transcutaneous electrical nerve stimulator Spikes, runaway pacemaker, ventricular fibrillation, non-specific
Peripheral nerve stimulator Spikes, loss of pacemaker spikes

The table includes data from references[5][7]

Identification

Artifacts can distort individual or all components (P, QRS, T waves and PR and ST segments) on the EKG. Most of the time, they are easily identifiable and neglected as they do not resemble any specific pattern. It is important to differentiate these artifactual changes from genuine changes to prevent misdiagnosis. If ST segments are affected by artifacts, either ST segment depression or elevation can occur on the EKG. These changes can be misinterpreted as myocardial ischemia or infarction.

Differentiating an Artifact from Ventricular tachycardia

Sometimes, EKG changes may mimic specific arrhythmias like ventricular tachycardia and atrial flutter or fibrillation.[8] It is important to differentiate these, as misdiagnosis can lead to inadvertent use of medications and procedures in such a patient[9][10].

  • Characteristics that can help in differentiate an artifact from ventricular tachycardia include[9]:
    • Absence of hemodynamic deterioration during the event.
    • Normal QRS complexes within the artifact.
    • An unstable baseline on the EKG before the event, after the event, or both.
    • Association with bodily movement.
  • Huang et al[10] also described 3 signs that may help in differentiating a tremor-induced pseudo-ventricular tachycardia from true ventricular tachycardia. Presence of any of these signs is suggestive of pseudo-ventricular tachycardia:
    • "Sinus" sign: One of the frontal leads (I, II and III) may present with sinus rhythm showing normal P, QRS, and T waves. The reason is that one of the upper limbs may be free off tremor.
    • "Spike" sign: Presence of regular or irregular tiny spikes among wide-QRS complexes.
    • "Notch" sign: Notches superimposed in the wide-QRS-like complex artifact, coinciding with the cycle length when sinus rhythm was recorded.

Electrode misplacement

Electrode misplacements are a common artifact that can mimic life-threatening arrhythmias. Early identification and replacement of electrodes can help in avoiding unnecessary therapies. An algorithm has been described previously[3], which may help in recognizing these artifacts.

REVERSE mnemonic: Approach to EKG artifacts[3]

EKG findings Explanation
R R wave is positive in lead aVR (P wave also positive) Reversal of left arm and right arm electrodes
E Extreme axis deviation: QRS axis between -90 and +180 Reversal of left arm and right arm electrodes
V Very low (<0.1 mV) voltage in an isolated limb lead Reversal of right leg and left arm or right arm electrodes
E Exchanged amplitude of P waves (P wave in lead I > lead II) Reversal of left arm and left leg electrodes
R R wave abnormal progression in precordial lead (pre-dominal R in V1 and S in V6) Reversal of precordial electrodes (V1 through V6)
S Suspect dextrocardia (negative P waves in lead I) Reversal of left arm and right arm electrodes
E Eliminate noise and interference (artifact mimicking tachycardias or ST-T changes

Table adapted from Baranchul et al's work[3]: Crit Care Nurse 2009;29:67-73 doi: 10.4037/ccn2009607

Other Common Artifacts

  • Electrodes on the torso: Placement of the electrodes on torso may lead to a change in vectors and produce pseudo-Q waves and pseudo-ST segment elevation, mimicking myocardial infarction.[3]
  • Telemetry interference: Superimposition of telemetry electrodes over the EKG electrodes or vice versa may cause ST segment deviation due to electromagnetic interference.[3]
  • Loose wire: Straight line may resemble systole and a wavy line may resemble fibrillation. However, it will be limited to one or two leads only.
  • Tall T wave: A tall T wave may be mistaken for an R wave and the digital heart rate would be higher than the actual pulse rate.
  • Lead placement: Obscuring of P waves may resemble a heart block.
  • Motion artifact: Chest percussions or physiotherapy may mimic ventricular fibrillation.

Consequences

Apart from the poor quality of EKG, artifacts can cause serious consequences particularly when they mimic genuine changes. If EKG artifacts are not recognized by physician, anesthesiologist or intensivist unneccasry diagnostic and therapeutic measures could be taken. Such actions may subject patients to invasive investigations or they may receive unnecessary medications like antiarrythmics.

Correction

  • Attention to basic principles such as proper electrodes placement and lead connections (as mentioned above) is required during EKG monitoring.
  • Well designed and maintained EKG measurement devices can withstand routine internal or external electrical and motion-related disturbances. However, it is not always possible to eliminate artifacts completely.
  • It is essential that physicians keep high vigilance and interpret EKG keeping artifacts in their differential diagnosis list.

Examples of Artifacts

The gallery of images below shows a variety of EKG artifacts that are due to tremor, movement, movement disorders and electrical interference. These artifacts should not be confused with an abnormality in the patient's electrocardiogram. Atrial fibrillation can be confused with these artifacts.




Video explaining ECG artifacts

{{#ev:youtube|LslIvYy2xUU}}

References

  1. Kligfield P, Gettes LS, Bailey JJ; et al. (2007). "Recommendations for the standardization and interpretation of the electrocardiogram: part I: The electrocardiogram and its technology: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society: endorsed by the International Society for Computerized Electrocardiology". Circulation. 115 (10): 1306–24. doi:10.1161/CIRCULATIONAHA.106.180200. PMID 17322457. Unknown parameter |month= ignored (help)
  2. Márquez MF, Colín L, Guevara M, Iturralde P, Hermosillo AG (2002). "Common electrocardiographic artifacts mimicking arrhythmias in ambulatory monitoring". Am. Heart J. 144 (2): 187–97. PMID 12177632. Unknown parameter |month= ignored (help)
  3. 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 Baranchuk A, Shaw C, Alanazi H; et al. (2009). "Electrocardiography pitfalls and artifacts: the 10 commandments". Crit Care Nurse. 29 (1): 67–73. doi:10.4037/ccn2009607. PMID 19182282. Unknown parameter |month= ignored (help)
  4. Genge JR (1991). "Contamination of breath methane samples in sterilized Vacutainer Tubes". Clin. Chem. 37 (11): 2019–20. PMID 1934491. Unknown parameter |month= ignored (help)
  5. 5.0 5.1 Patel S (2006). "Electrocardiographic artifact mimicking ventricular tachycardia during high-frequency oscillatory ventilation: a case report". Am. J. Crit. Care. 15 (3): 310–1. PMID 16632773. Unknown parameter |month= ignored (help)
  6. Sliwa JA, Marinko MS (1996). "Transcutaneous electrical nerve stimulator-induced electrocardiogram artifact. A brief report". Am J Phys Med Rehabil. 75 (4): 307–9. PMID 8777027.
  7. Patel SI, Souter MJ (2008). "Equipment-related electrocardiographic artifacts: causes, characteristics, consequences, and correction". Anesthesiology. 108 (1): 138–48. doi:10.1097/01.anes.0000296537.62905.25. PMID 18156892. Unknown parameter |month= ignored (help)
  8. Knight BP, Pelosi F, Michaud GF, Strickberger SA, Morady F (2001). "Physician interpretation of electrocardiographic artifact that mimics ventricular tachycardia". Am. J. Med. 110 (5): 335–8. PMID 11286946. Unknown parameter |month= ignored (help)
  9. 9.0 9.1 Knight BP, Pelosi F, Michaud GF, Strickberger SA, Morady F (1999). "Clinical consequences of electrocardiographic artifact mimicking ventricular tachycardia". N. Engl. J. Med. 341 (17): 1270–4. doi:10.1056/NEJM199910213411704. PMID 10528037. Unknown parameter |month= ignored (help)
  10. 10.0 10.1 Huang CY, Shan DE, Lai CH; et al. (2006). "An accurate electrocardiographic algorithm for differentiation of tremor-induced pseudo-ventricular tachycardia and true ventricular tachycardia". Int. J. Cardiol. 111 (1): 163–5. doi:10.1016/j.ijcard.2005.06.017. PMID 16026871. Unknown parameter |month= ignored (help)

Template:WH Template:WS