1. Wrap the wrist band around the sheath with the inflation port directed towards the fingers of the patient.  
1. Wrap the wrist band so that the green box marker is over the radial artery puncture site.  


[[File:Untitled10.jpeg|600px]]
[[File:Untitled10.jpeg|600px]]


2. Withdraw the radial sheath few centimeters so that the arm band can fit snugly and adjust the green box to align over the point of sheath insertion.
2. Withdraw the radial sheath few centimeters so that the wrist band can fit snugly and adjust the green box to align over the point of sheath insertion.


[[File:Radial compression device 3.jpeg|600px]]
[[File:Radial compression device 3.jpeg|600px]]
Line 55: Line 55:
[[File:Radial compression device 4.jpeg|200px]]
[[File:Radial compression device 4.jpeg|200px]]


4. Inspect the puncture site for oozing of blood. It's helpful to deflate the wrist band by 2 cc for a second, to visualize a little oozing to assure patent circulation of radial artery. Rapidly reinflate the wrist band to stop oozing.
4. Inspect the puncture site for oozing of blood. It's helpful to deflate the wrist band by 1 cc for a second, to visualize a little oozing to assure patent circulation of radial artery. Rapidly reinflate the wrist band to stop oozing.


[[File:Radial compression device 5.jpeg|300px]]
[[File:Radial compression device 5.jpeg|300px]]

Revision as of 00:24, 23 August 2014

Radial artery cathetarization Microchapters

Home

Patient Information

Overview

Radial Artery Anatomy

Advantages

Potential Pitfalls

Contraindications

Pre-procedure Assessment

Procedure

Achieving Hemostasis after Radial Artery Catheterization

Complications

Best Practices for Transradial Angiography

Case Studies

Case #1

Radial catheterization procedure On the Web

Most recent articles

Most cited articles

Review articles

CME Programs

slides

Images

American Roentgen Ray Society Images of Radial catheterization procedure

All Images
X-rays
Echo & Ultrasound
CT Images
MRI

Ongoing Trials at Clinical Trials.gov

US National Guidelines Clearinghouse

NICE Guidance

FDA on Radial catheterization procedure

CDC onRadial catheterization procedure

Radial catheterization procedure in the news

Blogs on Radial catheterization procedure

Directions to Hospitals Treating Radial artery cathetarization

Risk calculators and risk factors for Radial catheterization procedure

Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor-In-Chief: Varun Kumar, M.B.B.S.; Lakshmi Gopalakrishnan, M.B.B.S.

Overview

The transradial approach (TRA) for coronary/cardiac catheterization is slightly more challenging than the transfemoral route. It requires more maneuvers and catheter steering along the course of the radial artery to reach the level of the coronary sinus in the ascending aorta. The right transradial approach is usually preferred over the left side due to the ease of accessibility for the operating physician.

Catheter/GuideWire Selection

Catheter selection is an important key in the success of the procedure. Standard catheters (e.g Judkin's Left and Judkin’s right) can be used for the TRA. However due to the common anomalies that could be seen along the course of the radial, subclavien, and axillary arteries, special “dedicated Radial Catheters” had been developed (e.g. Tiger, Jacky, and Kimmy). The most commonly used guidewires are those with hydrophilic coating, due to the expected tortuosity of the arterial course.

Procedure

  1. Before the start of the procedure a “time out” should be taken to revise the indication, important labs, and the patency of the wrist circulation as discussed previously. Allen's test or Barbeau test.
  2. Make sure that all of the equipment are ready and recheck the sizes .
  3. The wrist should be shaved (if necessary) and cleansed in the usual sterile fashion. In addition, the groin should be prepped in case of access failure or the need for urgent IABP or a temporary venous pacemaker.
  4. Intravenous (IV) line should be started on the contralateral extremity. If an IV should be placed in the intervention extremity, it must be placed proximal to wrist preferably at the level of elbow.
  5. Sterilization of the whole arm should be done using the standard sterile techniques and covered in sterile drapes except for the region of the radial artery - distal part of the palmar surface of the forearm - which is left exposed for an easier access.
  6. Most operators prefer to give anxiolytics e.g Midazolam (1-2 mg) before the start of the procedure to decrease patient anxiety and hence decrease radial spasm.
  7. The arm should be fully extended; slightly elevated with full supination, and the wrist is hyperextended.
  8. Palpation of the radial artery against the styloid process of the radius with middle three fingers, to find the point of maximal impulse "PMI" is done.
  9. An amount of 0.5cc to 1.0 cc of local anesthetic is injected superficially over the PMI.
  10. Two techniques are used to puncture the radial artery to obtain access:-
    • Direct/single wall puncture : same used in femoral approach.
    • Transfixation/ double wall puncture: in which a special kind of puncture needle 18-21 guage is used to puncture the vessel. An indicator at the needle cap whould show successful penetration of the artery. Penetrate the other wall of the vessel then withdraw the needle leaving the canula inside the double puncture. Slowly withdraw the cannula untill a pulsitile flow of blood is noticed.
  11. An exchange length 0.035-0.038 inch J-tip guidewire is inserted.
  12. A small incision is made to the skin at the entry of the wire to fascilitate the sheath insertion.
  13. Radial sheath of 23cm long and 4-6Fr size should then be introduced.
  14. Through sidearm of the sheath, 5000U of heparin should be administered
  15. To reduce spasm, 500 micrograms of diltiazem can also be administered via the sidearm
  16. Coronary catheters are then advanced along the guidewire into aorta
  17. Left and right coronary arteries are then catheterized using Judkins, Amplatz or multipurpose catheter
  18. Hemostasis is achieved by direct pressure at the puncture site at the end of the procedure after removal of radial sheath
  19. Radial pulse, and writst circulation should be monitored after the procedure for several hours regularly.

Radial sheath kits are now available which contain hydrophilic coated sheaths in sizes 4-6Fr equipped with graduated introducers, various micropuncture needles and guidewire.


Radial Compression Devices

Radial Compression Devices are used for hemostasis after the removal of the radial sheath. In the picture below is one example of the devices used.

Steps

1. Wrap the wrist band so that the green box marker is over the radial artery puncture site.

2. Withdraw the radial sheath few centimeters so that the wrist band can fit snugly and adjust the green box to align over the point of sheath insertion.


3. Use the inflation syringe provided with the kit to inflate the wrist band by 10-15 cc of air. While inflating the band withdraw the sheath slowly and steadily using your other hand.

4. Inspect the puncture site for oozing of blood. It's helpful to deflate the wrist band by 1 cc for a second, to visualize a little oozing to assure patent circulation of radial artery. Rapidly reinflate the wrist band to stop oozing.


5. Continuously monitor radial artery pulsation distal to the wrist band, and check pulse oximeter for intact hand circulation.

6. Remove wrist band after 1-2 hours according to patient status.