You don't need to be Editor-In-Chief to add or edit content to WikiDoc. You can begin to add to or edit text on this WikiDoc page by clicking on the edit button at the top of this page. Next enter or edit the information that you would like to appear here. Once you are done editing, scroll down and click the Save page button at the bottom of the page.

Jump to: navigation, search

An optode or optrode is an optical sensor device that optically measures a specific substance usually with the aid of a chemical transducer.

Construction

An optode requires three components to function: a chemical that responds to an analyte, a polymer to immobilise the chemical transducer and instrumentation (optical fibre, light source, detector and other electronics). Optodes usually have the polymer matrix coated onto the tip of an optical fibre, but in the case of evanescent wave optodes the polymer is coated on a section of fibre that has been unsheathed.

Operation

Optodes can apply various optical measurement schemes such as reflection, absorption, evanescent wave, luminescence (fluorescence and phosphorescences), chemiluminescence, surface plasmon resonance. By far the most popular methodology is luminescence.

Luminescence in solution obeys the linear Stern-Volmer relationship. Fluorescence of a molecule is quenched by specific analytes e.g. Ruthenium complexes are quenched by oxygen. When a fluorophore is immobilised within a polymer matrix a miriad of micro-environments are created. The micro-environments reflect varying diffusion co-efficients for the analyte. This leads to a non-linear relationship between the fluorescence and the quencher (analyte). This relationship is modelled in various ways, the most popular model is the Two site model created by James Demas (University of Virginia).

The signal (fluorescence) to oxygen ratio is not linear, and an optode is most sensitive at low oxygen concentration, i.e. the sensitivity decreases as oxygen concentration increases. The optode sensors can however work in the whole region 0-100% oxygen saturation in water, and the calibration is done the same way as with the Clark type sensor. No oxygen is consumed and hence the sensor is stirring insensitive, but the signal will stabilize more quickly if the sensor is stirred after being put into the sample.

Popularity

Optical sensors are growing in popularity due to the low-cost, low power requirements and long term stability. They provide viable alternatives to electrode-based sensors or more complicated analytical instrumentation.

Major international conferences are devoted to their development e.g Europtrode VIII Tubeingen 2006, OFS 18, Cancun 2006.

See also

Oxygen sensorde:Optrode


Acknowledgement and Attribution Regarding Sources of Content

Some of the initial content on this page may be incorporated in part from copyleft sources in the public domain including wikis such as Wikipedia and AskDrWiki. Drug information for patients came from the The National Library of Medicine. Infectious disease information may have come from the Centers for Disease Control (CDC). Differential Diagnoses are drawn from clinicians as well as an amalgamation of 3 sources: 1.The Disease Database; 2. Kahan, Scott, Smith, Ellen G. In A Page: Signs and Symptoms. Malden, Massachusetts: Blackwell Publishing, 2004:3; 3. Sailer, Christian, Wasner, Susanne. Differential Diagnosis Pocket. Hermosa Beach, CA: Borm Bruckmeir Publishing LLC, 2002:7 .