K-space (MRI)

You don't need to be Editor-In-Chief to add or edit content to WikiDoc. You can begin to add to or edit text on this WikiDoc page by clicking on the edit button at the top of this page. Next enter or edit the information that you would like to appear here. Once you are done editing, scroll down and click the Save page button at the bottom of the page.

Jump to: navigation, search
Image:K-space partial Fourier.JPG
The conjugate symmetry of k-space
K-space is a formalism widely used in Magnetic Resonance Imaging.

Simply speaking, k-space is the temporary image space in which data from digitized MR signals are stored during data acquisition. When k-space is full (at the end of the scan), the data are mathematically processed to produce a final image. Thus k-space holds raw data before reconstruction.

K-space is in spatial frequency domain. Thus if we define kFE and kPE such that

k_{FE}=\bar{\gamma} G_{FE}m\Delta t

and

k_{PE}=\bar{\gamma} n\Delta G_{PE}

where FE refers to Frequency Encoding, PE to Phase Encoding, Δt is the sampling time (the reciprocal of sampling frequency), τ is the duration of GPE, \bar{\gamma} (Gamma bar) is 42MHz per Tesla, m is the sample number in the FE direction and n is the sample number in the PE direction (also known as partition number), the 2D-Fourier Transform of this encoded signal results in a representation of the spin density distribution in two dimensions. Thus position (x,y) and spatial frequency (kFE, kPE) constitute a Fourier transform pair.

K-space has the same number of rows and columns as the final image. During the scan, k-space is filled with raw data one line per TR.

Although a strict mathematical proof does not exist and counterexamples can be provided, in most cases it is safe to say that data in the middle of k-space contain the signal to noise and contrast information for the image, while data around the outside of the image contain all the information about the image resolution. This is the basis for advanced scanning techniques, such as the keyhole acquisition, in which a first complete k-space is acquired, and subsequent scans are performed by acquiring just the central part of the k-space; in this way, different contrast images can be acquired without the need of running full scans.

A nice symmetry property exist in the k-space, descending from the fact that the object imaged is a contrast-weighted proton density and thus a real quantity, relating the signal at two opposite locations in k-space:

S( − kFE, − kPE) = S * (kFE,kPE)

where the star denotes complex conjugation. Thus k-space information is somewhat redundant, and an image can be reconstructed using only one half of the k-space, either in the PE direction saving scan time (such a technique is known as half Fourier or half scan) or in the FE direction, allowing for lower sampling frequencies and/or shorter echo times (such a technique is known as half echo).

References

  • McRobbie D., et al. MRI, From picture to proton. 2003
  • Hashemi Ray, et al. MRI, The Basics 2ED. 2004.
WikiDoc Help Menu

Quick Start..

Editing basics

Advanced editing

Communicating your edits

Help Videos You Can Watch



Acknowledgement and Attribution Regarding Sources of Content

Some of the initial content on this page may be incorporated in part from copyleft sources in the public domain including wikis such as Wikipedia and AskDrWiki. Drug information for patients came from the The National Library of Medicine. Infectious disease information may have come from the Centers for Disease Control (CDC). Differential Diagnoses are drawn from clinicians as well as an amalgamation of 3 sources: 1.The Disease Database; 2. Kahan, Scott, Smith, Ellen G. In A Page: Signs and Symptoms. Malden, Massachusetts: Blackwell Publishing, 2004:3; 3. Sailer, Christian, Wasner, Susanne. Differential Diagnosis Pocket. Hermosa Beach, CA: Borm Bruckmeir Publishing LLC, 2002:7 .

Personal tools
related articles
viewed previously [ + ]